8: Pacing Troubleshooting
98
.
NOTE: Once the reason for the Pacing Stopped alarm has been resolved, that part of the alarm message is
removed from the display. The audio alarm continues. Press the
soft key to resume
pacing, remove the remainder of the alarm text from the display and silence the audio alarm.
WARNING: Observe the patient closely while pacing. Heart rate displays and alarms function during pacing but
can be unreliable. Do not rely on the indicated heart rate or heart rate alarms as a measure of the
patient’s perfusion status.
Troubleshooting
If your Efficia DFM100 does not operate as expected during pacing, see “Defibrillation and Pacing
Problems” on page 186.
Table 20 Pacing Alarms
Alarm Message Condition Type of Alarm Indication
Pacing has stopped. There has
been a power failure during
pacing.
High Priority
Non-Latching
Alarm
Red Alarm message
with alarm tone
Pacing has stopped. A pads off
condition has been detected
during pacing.
Pacing has stopped. The Efficia
DFM100 has detected an error
which prevents delivery of pacing
therapy.
Pacing has stopped. The Therapy
cable is disconnected from the
device.
Therapy is disabled due to an
equipment failure.
Pacing has stopped. The primary
ECG lead has become invalid in
Demand Mode pacing.
The actual delivered pace pulse
current is less than the selected
output.
The battery’s charge level is low.