Pulse Rate Alarms 10: Monitoring SpO
2
115
Changing SpO
2
Alarm Limits
To change the SpO
2
high and low alarm limits:
1 Press the Smart Select knob.
2 Turn the Smart Select knob to highlight
and press the Smart Select knob.
3 Select
and press the Smart Select knob.
4 Select
and press the Smart Select knob.
5 Turn the Smart Select knob to select the new high limit value then press the Smart Select knob.
6 Select the new low limit and press the Smart Select knob.
Enabling/Disabling SpO
2
Alarms
To e n a b l e / d i s a b le Sp O
2
alarms:
1 Press the Smart Select knob.
2 Turn the Smart Select knob to highlight
and press the Smart Select knob.
3 Select
and press the Smart Select knob.
4 Select
() and press the Smart Select knob.
WARNING: Turning alarms off prevents all alarms associated with the SpO
2
measurement from being
annunciated. If an alarm condition occurs, no alarm indication is announced.
Pulse Rate Alarms
You can turn on/off Pulse Rate alarms in all clinical modes where SpO
2
is available. The configured
alarm limits may be changed during use. Alarms are annunciated if measurements fall outside the
configured limits for high or low pulse rate. All Pulse Rate alarms are categorized as non-latching
alarms, meaning they are automatically removed when their alarm condition no longer exists.
Pulse Rate alarms can be generated for the conditions shown in Table 26. Once generated, they
appear as alarm messages in the Pulse status area right above the Pulse numeric. There are both audio
and visual alerts. For more information on alarms, see “Alarms” on page 36.
NOTE: Pulse alarms are off by default except when the Efficia DFM100 enters Pacing Mode.
Changing Pulse Rate Alarm Limits
To change the Pulse high and low alarm limits:
1 Press the Smart Select knob.
2 Turn the Smart Select knob to highlight
and press the Smart Select knob.
Table 26 Pulse Rate Alarms
Alarm Message Condition Type of Ala rm Indication
Pulse value is greater than the
Pulse high alarm limit.
Medium
Priority
Non-Latching
Alarm
Yellow alarm message
with audio tone
Pulse value is less than the Pulse
low alarm limit.