10: Monitoring SpO
2
Disabling SpO
2
Monitoring
116
3 Select  and press the Smart Select knob.
4 Select
 and press the Smart Select knob.
5 Turn the Smart Select knob to select the new high limit value then press the Smart Select knob.
6 Select the new low limit and press the Smart Select knob.
Enabling/Disabling Pulse Rate Alarms
To enable/disable Pulse alarms:
1 Press the Smart Select knob.
2 Turn the Smart Select knob to highlight
 and press the Smart Select knob.
3 Select
 and press the Smart Select knob.
4 Select
 () and press the Smart Select knob.
Disabling SpO
2
Monitoring
To d i s a b l e Sp O
2
monitoring:
1 Disconnect the sensor cable from the SpO
2
port. The message 
appears.
2 Select
 and press the Smart Select knob.
NOTE: If the sensor cable is disconnected accidentally, the message  appears.
Select
 and press the Smart Select knob. Secure the sensor connection to begin SpO
2
monitoring
again.
Caring for Sensors
Refer to the manufacturer’s instructions for care and cleaning of your sensors. To get the best results
from your reusable sensors, always handle the sensors and cable with care and protect them from
sharp objects. The sensor houses a sensitive electronic device that can be damaged. Harsh treatment
of the sensor reduces their useful life.
Troubleshooting
If your Efficia DFM100 does not operate as expected during SpO
2
Monitoring, see Table 52SpO
2
Problems” on page 188.