13: Data Management Event Summary
134
Logged when the Efficia DFM100 is evaluating the ECG rhythm.
(with settings)
Initial ECG bandwidth for display, printing and storage are logged.
Logged when the device detects a therapy equipment failure.
SpO
2
/ Logged when SpO
2
monitoring is connected and any subsequent
disconnect/connect.
(value, limit) Logged when the patient’s SpO
2
value is higher than the configured limit.
(value, limit) Logged when the patient’s SpO
2
value is lower than the configured limit.
(value, limit) Logged when the patient’s Desat value is lower than the configured limit.
Logged when the device is on and unable to acquire a Pleth waveform.
Logged when SpO
2
monitoring is on and the sensor is disconnected.
Logged when SpO
2
is on and a noisy signal is detected.
Logged when SpO
2
is on and light interference is detected at the SpO
2
sensor.
Logged when SpO
2
is on and a non-pulsatile SpO
2
sensor signal is detected.
Logged when SpO
2
is on and an SpO
2
equipment malfunction is detected.
Logged when SpO
2
is on and an erratic measurement condition occurs.
Logged when SpO
2
is on and the SpO
2
measurement update period exceeds 30
seconds.
Logged when SpO
2
is on and low perfusion occurs.
Logged when a non-critical SpO
2
error is detected.
/
Logged when SpO
2
alarms are turned on or off and any subsequent change.
(low) (high)
(limit)
Logged when SpO
2
alarms are turned on and when there is a change in alarm limits.
Pulse
(value, limit) Logged when the patient’s pulse is higher than the configured limit.
(value) (limit) Logged when the patient’s pulse is lower than the configured limit.
/
Logged when pulse alarms are turned on or off and any subsequent change.
(low limit) (high
limit)
Logged when pulse alarms are turned on and when there is a change in alarm limits.
EtCO
2
/ Logged when EtCO
2
monitoring is connected and any subsequent
disconnect/connect.
(value, limit) Logged when the patient’s EtCO
2
value is higher than the configured limit.
(value, limit) Logged when the patient’s EtCO
2
value is lower than the configured limit.
(type) Logged when a CO
2
sensor is connected to the Efficia DFM100.
Table 29 Event Information (Continued)
Logged Event Frequency