Event Summary 13: Data Management
135
Logged when the CO
2
sensor has not warmed up to operating temperature range.
Logged when the CO
2
sensor is unplugged or the filter line is disconnected.
Logged when the CO
2
sensor is unplugged.
/
Logged when CO
2
alarms are turned on or off and any subsequent change.
(low value) (high
value)
Logged when CO
2
alarms are turned on and when there is a change in alarm limits.
If using the Mainstream or Sidestream sensor, logged when the sensor is zeroing.
If using the Mainstream or Sidestream sensor, logged when the sensor needs to be
zeroed.
If using the Mainstream or Sidestream sensor, logged when the sensor zeroing is
complete.
(reason) If using the Mainstream or Sidestream sensor, logged when the sensor zeroing failed.
Logged when the CO
2
sensor has reached its end of life.
Logged when the CO
2
sensor needs servicing.
Logged when the CO
2
sensor detects a kinked or blocked filter line.
Logged when the CO
2
sensor detects the airway adapter is blocked.
Logged when the CO
2
sensor detects that it is over temperature.
Logged when the CO
2
sensor detects its result is out of range.
Logged when a non-critical EtCO
2
error is detected.
Logged when the Efficia DFM100 cannot communicate with the CO
2
sensor.
AwRR
/
Logged when AwRR alarms are turned on or off and any subsequent change.
(low value) (high
value)
Logged when AwRR alarms are turned on and when there is a change in alarm
limits.
(value, limit) Logged when the patient’s AwRR value is higher than the configured limit.
(value, limit) Logged when the patient’s AwRR value is lower than the configured limit.
(value) Logged when the Apnea alarm is displayed.
(time) Logged when there is a change to the Apnea alarm Limit setting.
NBP
Logged when a manual NBP measurement is requested.
(frequency value) Logged when an automatic NBP measurement is requested (including frequency
value).
Logged when an automatic NBP measurement is changed to manual (including
frequency value).
(frequency) Logged when the NBP frequency is changed.
/ Logged when NBP alarms are turned on or off and any subsequent change.
Table 29 Event Information (Continued)