Operational Check 15: Operational and Shift Check
159
Operational Check
Operational Checks supplement automated tests by verifying therapy cables, the ECG cable,
paddles, audio, charge and shock buttons, and the ability to deliver defibrillation and pacing therapy.
Operational Check also checks the SpO
2
, NBP and EtCO
2
modules and printer.
Operational Checks should be performed weekly to supplement the hourly, daily and weekly tests
the Efficia DFM100 performs automatically.
From Operational Check you can also print Automated Test and Operational Check results.
WARNING: When performing an Operational Check or Weekly Shock Test, confirm the Efficia DFM100 is not
connected to a patient.
NOTES: Do not run Operational Check with internal paddles attached. Perform a Weekly Shock Test to test
internal paddles.
To confirm the ECG cable used during an event is functioning properly, use that same cable during
Operational Check.
To enter Operational Check:
1 Turn the Therapy knob to Monitor.
2 Press the Smart Select knob.
3 Turn the Smart Select knob to highlight
 and press the Smart Select knob.
4 Select
 and press the Smart Select knob. See Figure 73 for Operational Check
menu options.
5 Confirm your exit from a clinical mode. Select Yes and press the Smart Select knob.
The following tests are performed during an Operational Check (see “Operational Check Tests and
Results on page 163):
Figure 73 Operational Check
General System Sync button SpO
2
Pads/Paddles ECG
•Therapy Knob •Therapy •EtCO
2
•Printer
•Charge button •Leads ECG •NBP •Battery
Leads ECG rerun (if necessary) Shock button Audio





Select  to run an Operational Check.
Select  to review a summary of
Operational Checks (see “Operational Check
Summaries on page 168).
Select  to review a summary of
Auto Tests (see Auto Test Summaries on
page 169).
Select  to exit the menu.