Operational Check 15: Operational and Shift Check
163
Table 44 Operational Check Tests and Results
Te s t
Device
Prompts
User Actions Results What to do if test doesn’t pass
General System -
tests internal
clock battery,
power supply and
internal memory
None None Pass - All tested systems are
functioning properly.
Fail - One or more of the
tested systems is not
functioning properly.
Take the device out of use and
call for service.
Therapy Knob -
tests if the
Therapy knob is
set to 170J and
functioning
properly
None Confirm the
Therapy knob is
at 170J.
Pass - The Therapy knob is
set to 170J and functioning
properly.
Fail - The Therapy knob is
not set to 170J.
Fail - The Therapy knob is
not functioning properly.
Confirm that the Therapy knob
is set to 170J and repeat
Operational Check. If the test
continues to fail, take the
device out of use and call for
service.
NOTE: When the Therapy knob fails, you restart and remain in Monitor Mode
regardless of the knob’s position on the dial.
Charge button -
tests the Charge
button’s
functionality
Depending on the cable
connected:
Pass - Charge button passed.
Fail - Charge button is not
functioning.
Fail - You used the Smart
Select knob to charge.
Repeat the test and make sure
you press the Charge button. If
Operational Check continues
to fail, take the device out of
use and call for service.
Therapy cable:
You are
prompted to
Move Therapy
knob to 150.
Confirm test
load/defibrillator
test plug is
attached and
press the Charge
button.
External Paddles:
You are
prompted to
Move Therapy
knob to 150.
Confirm the
paddles are
seated in their
pockets and
press the Charge
button.
If the device does not detect a
pressed Charge button within 10
seconds, you are prompted to use
the Smart Select knob to charge.