16: Maintenance Automated Tests
172
Automated Tests
The Efficia DFM100 performs many maintenance activities including three tests that run
automatically at regularly scheduled intervals when power is supplied and the device is off. The tests
assess operational performance and alert you if a problem exists.
Results of tests associated with critical device functionality are reported through the RFU indicator
and the Automated Test Summary Report. Results are also reported through statements on the
Efficia DFM100’s display when the device is turned on. Ta bl e 4 5 provides a brief explanation of the
tests and lists each tests frequency.
NOTE: If the DFM100 is turned on in the middle of an automated test, the test stops and the device starts
in normal operating mode.
Auto Test Summaries
You can review, print and export all Auto Test Summaries the Efficia DFM100 performs.
To view a summary of Automated Tests:
1 Turn the Therapy knob to Monitor.
2 Press the Smart Select knob.
3 Turn the Smart Select knob to highlight
 and press the Smart Select knob.
4 Select
 and press the Smart Select knob.
5 Select
 and press the Smart Select knob. The message 

appears.
6 Select
 if you wish to continue. Select  to return to Monitor Mode. Press the Smart Select
knob to confirm your choice.
7 If you selected
, the Efficia DFM100 displays the Automated Test Summary screen (see
Figure 80).
Table 45 Automated Tests
Tes t Frequency Description
Hourly Every hour Tests power supply, charge level of the battery, internal
communication across all critical modules and components
and also the device’s internal temperature.
Daily Daily after midnight
according to the
devices internal clock
Tests all hourly components as well as defibrillation, ECG,
pacing, SpO
2
, NBP and the printer.
Weekly Weekly, after midnight
Sunday morning
according to the
devices internal clock
Tests all daily components as well as various electrical
circuit tests and administers a 150J shock internally to test
the defibrillation circuitry.