16: Maintenance Battery Maintenance
176
Power-Related Alarms
Power-related alarms are generated for the conditions shown in Tabl e 4 8. Once generated, they
appear as alarm messages on the Efficia DFM100 display. There are both audio and visual alerts. For
more information on alarms, see “Alarms on page 36.
Table 48 Power-Related Alarms
Alarm Message Condition Type of Alarm Indication and
Location
 Battery power is low. High Priority if
pacing otherwise,
low priority
non-latching alarm
Red alarm message if
pacing, cyan if not
with audio tone in
Battery Status area.
 Battery power is critically low.
The device will shut down in
1 minute.
High Priority if
pacing otherwise,
medium priority
non-latching alarm
Red alarm message if
pacing, yellow if not
with audio tone in
Technical Alarm
area.
 Battery power is critically low.
The device is shutting down
now.
High Priority
Non-Latching
Red alarm message
with audio tone in
Technical Alarm
area.


A low voltage has been
detected.


Communications between the
device and the battery have
failed.
Medium Priority
Non-Latching
Yellow alarm message
with audio tone in
Battery Status area.
 The battery has reached its
end of life. Low Priority
Non-Latching
Cyan alarm message
with audio tone in
Technical Alarm
area.