17: Troubleshooting Symptoms
186


message on
display.
The waveform in Wave
Sector 1 is no longer valid
and another ECG source is
available.
Check that the monitoring electrodes/pads are properly
applied.
Use the Lead Select button to select another lead to
monitor.
Lead wire off message on the
display.
The specified monitoring
electrode is off or not
making proper contact with
the patient.
Check that the monitoring electrodes are properly
applied. If necessary, prepare the patients skin and
apply new electrodes.
 message
on the display.
Two or more limb lead
electrodes are off or not
making proper contact with
the patient.
Check that the limb lead electrodes are properly
applied.
This message does not display when V and RL are off
because the device assumes a 3-Lead cable is attached.
Table 50 ECG Problems (Continued)
Symptom Possible Cause Potential Solution
Table 51 Defibrillation and Pacing Problems
Symptom Possible Cause Potential Solution


message on
display.
A test load or defibrillator
test plug is attached to the
end of the Therapy Cable.
Remove the test load or defibrillator test plug and attach
multifunction electrode pads.
Pads impedance is less than
10 ohms.
Check pads connection with the patient.
 message is
displayed.
Therapy cable is not
attached. Pads/paddles
connection compromised.
Make sure the Therapy cable is connected and the
pads/paddles are making proper contact with the
patient.
The Shock button was not
pressed within the
configured time period.
No action required. If desired, charge the device and
press the Shock button.
 message
is displayed.
A shock has been aborted
due to high impedance.
Check paddles connection with the patient. Remove
paste, moisture, or any other conductive material
between the paddles and patient.
 message is
displayed.
Check pads connection with the patient.

message is displayed.
The pads or paddles Therapy
cable is not connected to the
device.
Confirm all cables are properly connected.

message is displayed.
The pads cable is not
connected to the Therapy
cable or pads are not
properly applied to the
patient.
Confirm all cables are properly connected and pads are
correctly placed on the patient.
 message
is displayed.
An energy greater than 50J
was selected with internal
paddles connected.
Re-set energy setting to 50J or less.