19: Specifications Specifications
200
The delivered energy accuracy is ±10% or ±1J whichever is greater for all energy settings.
Charge times:
Less than 5 seconds to the recommended adult energy level (150 Joules) with a new fully-charged
battery installed.
Less than 6 seconds to the selected energy level (up to 200 Joules) with a new fully-charged battery
installed, even after the delivery of 15 discharges at maximum energy.
Less than 15 seconds to the selected energy level while connected to AC power only, even when
operating on 90% of the rated mains voltage.
The device powers on in manual defibrillation mode ready to deliver shock in less than:
• 23 seconds with AC power only and at 90% of rated mains voltage.
• 15 seconds with a new, fully-charged battery even after 15 discharges of maximum energy.
Table 58 Delivered Energy Accuracy
Nominal Delivered Energy vs. Load Impedance
Selected
Energy
Load Impedance (ohms) ±2%
25 50 75 100 125 150 175
1 J 1.2 1.3 1.3 1.2 1.1 1.0 0.9
2 J 1.7 2.0 2.1 2.0 1.9 1.7 1.6
3 J 2.6 3.0 3.1 3.2 3.2 3.1 2.9
4 J 3.5 4.0 4.2 4.3 4.4 4.5 4.3
5 J 4.3 5.0 5.2 5.4 5.5 5.6 5.4
6 J 5.2 6.0 6.3 6.5 6.6 6.7 6.5
7 J 6.1 7.0 7.3 7.6 7.8 7.8 7.6
8 J 6.9 8.0 8.4 8.6 8.9 8.9 8.7
9 J 7.8 9.0 9.4 9.7 10 10 9.8
10 J 8.7 10 10 11 11 11 11
15 J 131516 161717 16
20 J 172021 222222 22
30 J 263031 323333 33
50 J 435052 545556 54
70 J 617073 767878 76
100 J 87 100 105 108 111 111 108
120 J 104 120 126 130 133 134 130
150 J 130 150 157 162 166 167 163
170 J 147 170 178 184 188 189 184
200 J 173 200 209 216 222 223 217