Time from the initiation of analysis in AED mode until ready to deliver shock is less than 23 seconds
with:
AC power only and at 90% of rated mains voltage.
A new, fully charged battery even after 15 discharges of maximum energy.
The device powers on in AED mode ready to deliver shock in less than:
32 seconds with AC power only and at 90% of rated mains voltage.
24 seconds with a new, fully charged battery even after 15 discharges of maximum energy.
Patient Impedance Range: Minimum: 25 ohm (external defibrillation); 15 ohm (internal
defibrillation); Maximum: 250 ohm. Actual functional range may exceed these values.
Manual Defibrillation Mode
Manual Output Energy (Selected): 1-10, 15, 20, 30, 50, 70, 100, 120, 150, 170, 200 Joules;
maximum energy limited to 50J with internal paddles.
Controls: On/Off Therapy knob, Charge, Shock, Sync, ECG Lead Select, Patient Selection, Print,
Mark Events, Reports, Alarms, Smart Select knob.
Energy Selection: Front panel Therapy knob.
Charge Control: Front panel button; button on external paddles.
Shock Control: Front panel button; buttons on external or switched internal paddles.
Synchronized Control: Front panel Sync button.
Synchronized Shock Timing: Maximum time from R-Wave detected to shock delivered is 25ms, as
measured with oscilloscope from peak of input QRS wave to leading edge of defibrillation discharge
into a 50 ohm test load.
Indicators: Text prompts, audio alerts, QRS beeper, battery status, Ready For Use (RFU), External
Power, Sync Mode.
Armed Indicators: Charging/charged tones, flashing shock button on front of panel and on external
paddles, energy level indicated on the display.
Figure 81 Smart Biphasic Waveform
0
500
1000
1500
2000
2500
-150 0
-100 0
-500
0 5 10 15 20
Philips Smart Biphasic Waveform at 200J into 25-175 Ohms
Voltage (V)
Time (ms)