Specifications 19: Specifications
207
Wavelength Range: 500-1000 nm (Information about wavelength range can be useful to users,
especially those performing photodynamic therapy.)
Desat Alarm Signal Generation Delay: 20 sec
Pulse Rate Measurement Range: 30-300 bpm
Pulse Rate Resolution: 1 bpm
Pulse Rate Accuracy: ±2% or 1 bpm whichever is greater
Pulse Response Time (90 to 120 bpm): average 18.0 seconds, standard deviation 0.86 seconds
Pulse Alarm Range:
Low Limit: 30-295 (Adult and Infant/Child)
High Limit: 35-300 (Adult and Infant/Child)
EtCO
2
Weight: Mainstream: 78 g (2.75 oz.); Sidestream: 272 g (9.6 oz.)
Dimensions: Mainstream: 43 mm (W) x 33 mm (H) x 23 mm (L); 1.69 in (W) x 1.29 (H) x .90 in
(L); Sidestream: 66 mm (W) x 38 mm (H) x 89 mm (L); 2.6 in (W) x 1.5 in (H) x 3.5 in (L)
Range: 0-150 mmHg
Resolution: 1 mmHg (0.1 kPa)
Accuracy: 0 - 40 mmHg ± 2 mmHg; 41 - 70 mmHg ± 5% of reading; 71 - 100 mmHg ± 8% of
reading; 101 - 150 mmHg ± 10% of reading. Gas at 25C.
Drift of Measurement Accuracy: Over any 24 hour period, the specified measurement accuracy is
maintained.
Warm-up time: 2 minutes at 25°C.
System Response Time: Sidestream: 3.5 seconds typical.
Alarm Delay Time: (after alarm condition has been met) Mainstream - less than 5 sec; Sidestream -
less than 8 sec; Measurement Method: Peak EtCO
2
value within a 10 sec window.
Sample Flow Rate: Sidestream - 50 ml/min ±10ml
Alarm Range:
Low Limit: 10-140 mmHg (Adult, Infant/Child)
High Limit: 20-145 mmHg (Adult, Infant/Child)
AwRR
Range: 0-150 rpm
Resolution: 1 rpm
Accuracy: ±1 rpm
Alarm Range:
Low Limit: 0-99 rpm (Adult, Infant/Child)
High Limit: 10-100 rpm (Adult, Infant/Child)
Alarm Delay Time: (after alarm condition has been met) Mainstream - less than 5 sec; Sidestream -
less than 8 sec; Measurement Method: AwRR - based on the last 8 detected breaths; Apnea -
Following the configured Apnea delay time.