19: Specifications Specifications
210
Other considerations:
• The Efficia DFM100 is not suitable for use in the presence of concentrated oxygen or a
flammable anesthetic mixture with air, oxygen or nitrous oxide.
• Hazards arising from software errors were minimized by the product’s compliance with the
software requirements contained in ISO 62304/YY/T0664.
Mode of Operation: Continuous
AC Line Powered: 100-240 VAC, 50 or 60 Hz, 1 - 0.46 A, Class I Equipment
Battery Powered: Minimum 14.4 V, Rechargeable Lithium Ion
Hazardous Waste:
USB Device
Correct Drive: Use the Philips USB Drive that came with your device or is orderable under part
number 989803171261.
Pb Hg Cd Cr6+ PBB PBDE
•
OOOOO
• = more than one of the device’s raw material has this harmful substances and concentration
over than standard concentration limit.
O = all the raw material concentrations of the device within allowed limits.