3: Operating the Device The Display
30
Status Area
The Status Area contains key device and patient information, see Figure 27.
Patient information includes:
• Name/ID – If entered, displays the patient’s name, patient’s ID or nothing depending on
the current configuration. See Table 32 “General Settings” on page 149.
• Date and Time – The current date and time
• Paced Status – Internally paced or non-paced. The area remains blank if paced status has
not been selected. (This information is not displayed in Pacing and AED modes.)
• Event Timer – The elapsed time for the current event, displayed in hours:minutes:seconds
• Patient Category and Weight Range – Adult (defined as ≥ 25kg) or Infant/Child (defined as
< 25kg). The background color changes with the selected patient category for Adult
(orange) and Infant/Child (blue).
Device information includes:
• Clinical Alarm Indicator - Indicates the state of global alarms. If alarms are on and audio is
not paused, the indicator is blank. If they are off, the indicator says
and uses the
Alarms Off icon .
• Technical Alarms - Technical alarms not related to the battery or pacing are displayed in the
Technical Alarms area. A triangle indicates there are multiple alarms present, alternating
on the display.
• Battery Status – If a battery is installed, the battery icon indicates the level of charge
remaining. See Figure 28.
Figure 27 Status Area
Figure 28 Battery Charge Level
Event Timer
Paced
Status
Technical Alarm AreaClinical Alarm Area
Battery
Status
Date and
Time
Multiple Technical
Alarms Indicator
Patient
Name/ID
Patient Category
Weight Range
No
Battery
< 40%
remaining
≥ 40%− < 65%
remaining
≥ 65%− < 90%
remaining
≥ 90%
remaining
Low
Battery
Condition*
* - See “Low Battery Conditions” on page 27.