Alarms 3: Operating the Device
37
Clinical Mode Alarm Notification
When in a Clinical Mode, the Efficia DFM100 can be configured to react differently when an alarm
condition occurs. See Tab le 3 .
All alarm conditions are cleared when you switch from a clinical mode to a non-clinical mode.
If you intentionally disconnect a sensor, an alarm sounds. Press the Smart Select knob to stop the
alarm. The Efficia DFM100 prompts you to confirm your selection. Press the Smart Select knob
again.
WARNINGS: Silencing either audio or audio and visual indications of active alarms can result in missed alarm
conditions and also inhibit indications of new alarm conditions.
Confirm alarm limits are appropriate for the patient each time there is a new patient event.
Do not set alarm limits to such extreme values that render the alarm system useless.
A potential hazard exists if different alarm limits are used for the same or similar equipment in any
single area.
Table 3 Alarm Notification Types
Type Alarm Location
See Alarm Notification
Locations
on page 38.
Alarm Status Comments
Alarms
On
None None Both visual and audio
indications are on.
All alarms are on.
Alarm
Audio
Pause
(message includes duration of the pause)
Upper left corner of
the display
underneath the
Patient Category.
Only visual indications
are on for the
configured Audio Pause
timeframe. Audio
returns when the pause
timeframe is complete.
This message appears
when you press the
Alarms button and the
configured alarm pause is
set to something other
than indefinite.
Alarm
Audio
Off
Upper left corner of
the display
underneath the
Patient Category.
Only visual indications
are on.
This message appears
when you press the
Alarms button and the
configured alarm pause is
indefinite.
In the Menu area
above the Smart
Select knob.
When displayed, press
the Smart Select knob to
turn Audio off.
Alarm
Audio
Pause
In the Menu area
above the Smart
Select knob.
Only visual indications
are on for the
configured Audio Pause
timeframe. Audio
returns when the pause
timeframe is complete.
When displayed, press
the Smart Select knob to
pause Audio.
Alarms
Off
When in AED
Mode, upper left
corner of the display
underneath the
Patient Category.
Both audio and visual
alarm indications are
off.
Press the Alarms button
to activate alarms in AED
mode.




