4: ECG Monitoring Preparing to Monitor ECG
52
CAUTIONS: Beware of patient cables, including ECG monitoring equipment when used with high frequency
surgical equipment.
Line isolation monitor equipment may produce power line transients that may resemble actual
cardiac waveforms and thus inhibit heart rate alarms. Keep electrode wires and cables away from
power cords to minimize this problem.
Conductive parts of electrodes and associated connectors for applied parts, including the neutral
electrode, should not contact other conductive parts including earth.
NOTES: Signals from TENS or ESU units can cause artifact.
For patients who exhibit intrinsic rhythm only, the risk of missing cardiac arrest may be reduced by
monitoring these patients with low heart rate limit at or slightly above the basic/demand pacemaker
rate. A low heart rate alarm alerts you when the patient begins pacing. Proper detection and
classification of the paced rhythm can then be determined.
When using the ECG Analog Output, detected internal pacemaker pulses are inserted into the
output as pulses of less than 3.5 ms in width when measured at 50% of the peak of the pulse. The
amplitude of the inserted pulse is >40% and <70% of the detected pacemaker pulse amplitude for
pulses widths of 0.5 ms to 2.0 ms.