5: AED Mode Option Using AED Mode to Defibrillate
72
“Press pads firmly to patient’s
bare skin. Pads must not be
touching clothing or each other.
If needed, remove hair from
patient’s chest.
”
The Therapy cable is
connected. A pads off
condition still exists.
Securely connect the
Therapy cable and the
pads connector. See
“Connecting
Multifunction
Electrode Pads.” on
page 8.
Confirm there is a
good pads contact with
the skin.
“
Be sure pads connector is
completely inserted.
”
“Poor pads contact, Replace
pads.
”
Replace the pads.
“
Begin CPR.”
Begin CPR, if
indicated.
Table 12 AED Mode Pads Off Messages (Continued)
These messages may appear when the multifunction electrode pads are not secured to the patient.
Audio Message
L = long prompts S = short prompts
Not noted = Both
Display Text Condition User Action
Table 13 AED Mode Marginal Impedance Messages
These messages may appear when the Efficia DFM100 detects a higher pad impedance than expected in the average
patient. The cause could be excessive hair on the chest, expired or dried out pads, or poor pad placement/contact on
the patient.
Audio Message Display Text Condition User Action
“
Press pads firmly to patient’s
bare skin.
”
When the device is not
charged, charging or
delivering a shock or in a
pause period, pads
impedance condition is
marginal.
Recheck pads.
Confirm they are
securely connected to
the patient.
“
Press pads firmly to patient’s
bare skin. Pads must not be
touching clothing or each other.
If needed remove hair from
patient’s chest.
”
“
Poor pads contact, Replace
pads.
”
Recheck pads.
Confirm they are
securely connected to
the patient. If they are,
remove them and
replace with new pads.
“
Begin CPR.” None or
Good pads contact
cannot be established.
If CPR is indicated,
begin CPR.