5 Measurement Settings
38
1
The availability of these settings is depending on the Alarms from setting.
2
This setting is only visible when Mode is set to Auto.
3
These settings are only visible when Mode is set to Sequence, and you select Setup Sequence in the
Setup NBP menu.
4
Only available for NBP measurements with a cableless NBP pod.
NBP Configuration Implications
Alarms from
Your monitor must be configured to Alarm Mode > All to enable alarms in general.
The
Alarms from setting provides you a list of selections to decide whether alarms are derived from the
systolic, diastolic, or mean NBP. Choices are:
•
Sys.
• Dia.
• Mean
• Sys & Dia
• Dia & Mean
• Sys & Mean
• Sys&Dia&Mean
The chosen combination is displayed directly under the NBP label on the Screen as following:
Depending on the chosen
Alarm from setting, for an NBP value of for example 139/103, the following
alarms will be issued:
Measurement Time C Timestamp Time to Next
Aging Time C 10 min 1 min, 2 min, 2.5 min, 3 min, 5 min, 10 min,
15 min, 20 min, 30 min, 45 min, 1 h, 2 h
Automatic Start
4
C Off Off, Slow, Fast
Color C Red Red, Green, Yellow, Blue, Magenta, Cyan,
White, Pink, Orange, Light Green, Light Red
Setting Item Mode Factory Default Choice/Range
Alarms from Issued Alarm
Sys. No alarm issued
Dia. NBPd High
Mean NBPm High
Sys & Dia NBPd High