19 Maintenance Operational Check
228
Pads/Paddles ECG Checks ECG acquisition
through pad/paddles.
None, if test passes. If test fails the
following prompt is displayed at
the end of all remaining tests:
Pads/Paddles ECG Test failed
with cable. Disconnect therapy
cable to rerun test without cable.
If the Pads/Paddles test fails with
the cable and passes without the
cable, the cable is bad. Replace
the Therapy cable and rerun
Operational Check.
If the Pads/Paddles test fails with
and without the cable, see
“Troubleshooting” on page 249.
Battery A
Battery B
Checks the total battery
capacity (combination of both
batteries) and calibration status
of the batteries in
Compartments A and B.
None. None.
SpO
2
Checks the internal SpO
2
PCA.
The SpO
2
cable is not tested.
None. None.
NBP Checks to see if the NBP
module is functioning;
determines if it is due for
calibration.
None. None.
CO
2
Checks to see if the CO
2
module
is functioning;
determines if it is due for
calibration.
None. None.
Invasive Pressures Checks to see if the invasive
pressure hardware is working
properly.
None. None.
Temperature Checks to see if the temperature
hardware is working properly.
None. None.
Bluetooth Checks for the presence of the
Bluetooth card and database
integrity.
None. None.
Printer Runs a printer self test. None. None.
Table 57 Operational Checks Test (Continued)
Tes t Description Prompts Action