EasyManua.ls Logo

Philips HeartStart OnSite M5066A - After Using the HeartStart Onsite; After Each Use

Philips HeartStart OnSite M5066A
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PHILIPS MEDICAL SYSTEMS
4-1
4
4 AFTER USING THE HEARTSTART ONSITE
AFTER EACH USE
1. Check the outside of the OnSite for signs of damage, dirt, or
contamination. If you see signs of damage, contact Philips for technical
support. If the OnSite is dirty or contaminated, clean it according to the
guidelines in Chapter 5, “Maintaining the HeartStart OnSite.
2. The single-use pads must be replaced after being
used. Locate the latch at the top edge of the
OnSite and slide it to the side. The pads cartridge
will be released. Lift out the used pads cartridge.
3. Remove a new SMART Pads Cartridge from its
package and insert the cartridge into the cartridge
well on the front of the OnSite. It should click into place when properly
seated. The green PULL handle should be all the way down.
NOTE: To prevent the pads’ adhesive gel from drying out, do not open the
hard cover or film seal of the cartridge until you need to use the pads.
4. Check supplies and accessories for damage and expiration dates. Replace
any used, damaged or expired items. Use a new maintenance tag to record
the pads expiration date for the new installed pads cartridge. If you replace
the spare pads and/or battery be sure to record the dates for them on the
maintenance tag as described in Chapter 2. Then sign and date the
inspection log/maintenance booklet.
55+ lb
s / 25+ k g
55+ l bs / 25+ kg