EasyManua.ls Logo

Philips HeartStart OnSite M5066A - Maintaining the HeartStart Onsite; Routine Maintenance; Periodic Checks

Philips HeartStart OnSite M5066A
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PHILIPS MEDICAL SYSTEMS
5-1
5
5 MAINTAINING THE HEARTSTART ONSITE
ROUTINE MAINTENANCE
The OnSite is very simple to maintain. The OnSite performs a self-test every
day. In addition, a battery insertion self-test is run whenever a battery is installed
in the device. The OnSite’s extensive automatic self-test features eliminate the
need for any manual calibration. The OnSite has no user-serviceable parts.
WARNING: Electrical shock hazard. Do not open the OnSite, remove its
covers, or attempt repair. There are no user-serviceable components in the
OnSite. If repair is required, return the OnSite to Philips for service.
REMINDERS:
Do not leave the OnSite without a pads cartridge installed; the OnSite will
start chirping and the i-button will start flashing. For directions on changing
the pads cartridge, see Chapter 2, “Setting up the HeartStart OnSite.
The OnSite runs daily self-tests. As long as the green Ready light is blinking,
it is not necessary to test the OnSite by initiating a battery insertion self-
test. This uses battery power and risks draining the battery prematurely.
PERIODIC CHECK S
Other than the checks recommended after each use of the OnSite, maintenance
is limited to periodically checking the following:
Check the green Ready light. If the green Ready light is not blinking, see
Troubleshooting Tips, below.
Replace any used, damaged or expired supplies and accessories
Check the outside of the OnSite. If you see cracks or other signs of
damage, contact Philips for technical support.
Record each periodic check in your inspection log/maintenance booklet.

Table of Contents

Related product manuals