PHILIPS MEDICAL SYSTEMS
E-1
E
E TECHNICAL INFORMATION
HEARTSTART ONSITE DEFIBRILLATOR SPECIFICATIONS
The specifications provided in the following tables are nominal values. Additional
information can be found in the Technical Reference Manuals for HeartStart
Automated External Defibrillators, located online at www.philips.com/
productdocs.
PHYSICAL
ENVIRONMENTAL
category specifications
size 7.2cm H x 19cm D x 21cm W (2.85” H x 7.40” D x 8.30” W).
weight Approximately 1.5 kg (3.3 lbs) with battery and pads cartridge installed.
category
specifications
temperature and
relative humidity
Operating (battery and pads cartridge installed):
0° to 50° C (32° to 122° F);
0% to 95% RH (non-condensing).
Standby (between uses with battery and pads cartridge installed):
10° to 43° C (50° to 109° F);
10% to 75% RH (non-condensing).
Storage/shipping (with battery and pads cartridge):
-20° to 60° C (-4° to 140° F) for up to 2 days;
0% to 85% RH (non-condensing).
altitude Operates at 0 to 4,572 m (15,000 feet); can be stored at up to 2,591 m
(8,500 feet), in standby mode.
atmospheric pressure Operates at 1013 hPa to 590 hPa; can be stored at up to 750 hPa, in standby
mode.
shock/drop abuse
tolerance
Withstands 1 meter (3.3 foot) drop to any edge, corner, or surface.