PHILIPS MEDICAL SYSTEMS
E-2
CONTROLS AND INDICATORS
vibration Operating: meets EN1789 random, road ambulance.
Standby: meets EN1789 swept sine, road ambulance.
sealing Meets IEC 60529 class IP21.
Protected against access to hazardous parts with a finger and protected
against ingress of solid foreign objects of 1.25 cm (0.5 in) diameter and
greater per IEC 60529 class IP2x.
Protected against a uniform flow of water drops over the defibrillator per IEC
60529 class IPxI.
ESD/EMI (radiated and
immunity)
See Electromagnetic Conformity tables.
category
specifications
category specifications
controls Green SMART Pads Cartridge handle
Green On/Off button
i-button (flashes blue)
Orange Shock button
indicators Ready light: green; blinks when the OnSite is in standby mode (ready for use);
solid when the defibrillator is being used.
i-button: flashes blue when information is available, on solid during patient
care pause.
Caution light: flashes when the OnSite is analyzing, comes on solid when the
OnSite is ready to deliver a shock.
Shock button: orange, flashes when the OnSite is charged and ready to
deliver a shock.
audio speaker Provides voice prompts and warning tones during normal use.
beeper Provides chirps when troubleshooting is needed.