EasyManua.ls Logo

Philips HeartStart OnSite M5066A - Yes, No; Yes

Philips HeartStart OnSite M5066A
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
F-5
PHILIPS MEDICAL SYSTEMS
F
CPR Coaching
adult ventilation
instruction
Yes, No Yes Optional CPR Coaching includes
rescue breaths at the rate
determined by the CPR Coaching
compression:ventilation ratio for
adults when an adult pads cartridge
is installed.
NOTE: if this parameter is configured to
NO, CPR Coaching will always be
compressions-only when an adult pads
cartridge is installed.
CPR Coaching
infant/child ventilation
instruction
Yes, No Yes Optional CPR Coaching includes
rescue breaths at the rate
determined by the CPR Coaching
compression:ventilation ratio for
infants and children when an infant/
child pads cartridge is installed.
NOTE: if this parameter is configured to
NO, CPR Coaching will always be
compressions-only when an infant/child
pads cartridge is installed.
CPR Coaching
compression:ventilation
ratio
30:2 adult and
30:2 infant/child
30:2 adult and
15:2 infant/child
15:2 adult and
15:2 infant/child
30:2 adult and
30:2 infant/
child
If the user presses the i-button for
optional CPR Coaching during a
protocol pause or NSA pause, the
OnSite provides coaching in basic
CPR for cycles of 30 compressions
and 2 ventilations for adults,
children, and infants. Pauses begin
and end with compressions.
parameter settings default default description