M4735A Service Manual 4-47
Internal Assemblies - Top Case
4
After Repair
After the repair is complete, perform the following steps.
1. Replace the Shield Plate.
See "Shield Plate" on page 4-40.
2. Replace and reconnect the Control PCA.
See "Control PCA" on page 4-36.
3. Reassemble the case.
See "Closing the Case" on page 4-99.
4. Restore the customer’s configuration.
Enter Configuration Mode and reload the customer’s configuration from
the data card, or reconfigure the unit, as needed. See "Configuration
Mode" on page 2-11.
5. Test performance.
Conduct Performance Verification Testing as described in "Performance
Verification and Safety Tests" starting on page 2-1.