Internal Assemblies - Top Case
4-78 Removal and Replacement
2. Restore the customer’s configuration.
Enter Configuration Mode and reload the customers configuration from
the data card, or reconfigure the unit, as needed. See "Configuration
Mode" on page 2-11.
3. Test performance.
Conduct Performance Verification Testing as described in "Performance
Verification and Safety Tests" starting on page 2-1.