Cleaning the Ergonomic Grip
1. Wipe the grip with a dry or water-moistened soft cloth. You can use soapy
water, a 70% isopropyl alcohol solution (rubbing alcohol), or a 10% bleach
solution.
2. Remove any cleaning solution residue with a water-moistened soft cloth. Do
not allow cleaning solutions to air dry on the grip.
Transducer Storage
Use the appropriate guidelines for storing transducers for transport, and daily
and long-term storage.
Storage for Transport
If a carrying case is provided with your transducer, always use the carrying case
to transport the transducer from one site to another. Follow these guidelines
to properly store transducers for transport:
• Make sure that the transducer is clean and disinfected before placing it in
the case to avoid contaminating the foam that lines the carrying case.
• Place the transducer in the case carefully to prevent kinking of the cable.
• Before closing the lid, make sure no part of the transducer is protruding
from the case.
• Wrap the case in plastic material containing air pockets (bubble wrap), and
pack the wrapped case in a cardboard carton.
• To avoid damaging the shaft or steering mechanism of TEE transducers, do
not bend or coil the flexible shaft of the transducer in less than a 0.30-m
(1-ft) diameter circle.