Cleaning the Ergonomic Grip
1. Wipe the grip with a dry or water-moistened soft cloth. You can use soapy
water, a 70% isopropyl alcohol solution (rubbing alcohol), or a 10% bleach
solution.
2. Remove any cleaning solution residue with a water-moistened soft cloth. Do
not allow cleaning solutions to air dry on the grip.
Transducer Storage
Use the appropriate guidelines for storing transducers for transport, and daily
and long-term storage.
Storage for Transport
If a carrying case is provided with your transducer, always use the carrying case
to transport the transducer from one site to another. Follow these guidelines
to properly store transducers for transport:
Make sure that the transducer is clean and disinfected before placing it in
the case to avoid contaminating the foam that lines the carrying case.
Place the transducer in the case carefully to prevent kinking of the cable.
Before closing the lid, make sure no part of the transducer is protruding
from the case.
Wrap the case in plastic material containing air pockets (bubble wrap), and
pack the wrapped case in a cardboard carton.
To avoid damaging the shaft or steering mechanism of TEE transducers, do
not bend or coil the flexible shaft of the transducer in less than a 0.30-m
(1-ft) diameter circle.
iE33 User Manual
220
4535 616 23901