Key to Test 2 Diagram
Stray capacitance from the ultrasound
system power wiring to the system
grounded metal chassis (1 to 3 MΩ
reactance)
C
Impedance between the metal parts of the
transducer and a test electrode placed in
the bucket of saline solution (about 850 kΩ
with an intact outer insulating layer, 500
with a hole in the layer)
Z
Microammeter to measure leakage current
from a line supply to the transducer and
back to earth ground through electrode Z
and the equipment chassis
A
Line power source, either 110 Vac or
220 Vac
e
Current caused by e and stray capacitance,
and optionally Z
I
Open earth lift ground switchS1
Line polarity switchS2
Isolation test buttonS3
Line mains supplyL
Neutral mains supplyN
Earth groundE
233
iE33 User Manual
4535 616 23901
9
Intraoperative Transducers