If a system message indicates a calibration failure, you will be unable to use the
transducer until you perform the following procedure.
1. Ensure that the transducer shaft and tip are straight (the neutral position
indicators must be aligned with the array rotation buttons).
2. Turn the locking lever on the transducer connector counterclockwise and
pull the connector out of the receptacle.
3. Reseat the connector in the receptacle and turn the locking lever clockwise.
4. Select the transducer and preset.
5. If you still see the calibration-failure message, contact your Philips service
representative.
S7-3t TEE Transducer Description
The S7-3t TEE transducer is described below.