3. Once the tip is positioned properly, lock the deflection control knob by sliding
the deflection lock so that the indicator is red.
Manipulating the S7-3t Transducer Tip
Deflection lock1
Deflection control2
S7-3t TEE Array Rotation
You can rotate the imaging array continuously from 0 degrees (transversal plane)
through 90 degrees (longitudinal plane) to 180 degrees (transversal plane, left/right
inverted). Because the transducer allows you to select all planes between 0
degrees and 180 degrees, you can scan a conical imaging volume.
iE33 User Manual
270
4535 616 23901
Transesophageal Transducers
10