CAUTIONS
• Using non-recommended disinfectants, using incorrect solution strengths,
or immersing a transducer deeper or longer than recommended can damage
or discolor the transducer and voids the transducer warranty.
• Do not immerse transducers longer than the minimum time needed for
your level of disinfection. For information on the levels of disinfection
requirements, see "Choosing a Disinfectant" on page 301.
Disinfecting Transducers by Immersion
Before performing this procedure, read the warnings and cautions in "Disinfection
of Transducers by Immersion (High-Level Disinfection)" on page 309.
Methods for disinfecting transesophageal (TEE) and non-TEE transducers differ.
For TEE transducers, see "Disinfecting TEE Transducers by Immersion" on page
311 or "Disinfecting TEE Transducers in an Automated Disinfector" on page 315.
1. Clean the transducer according to "General Cleaning for All Transducers"
on page 302.
2. Mix the disinfection solution compatible with your transducer (see
"Disinfectants and Cleaning Solutions Compatibility Table" on page 322)
according to label instructions for solution strength.
3. Immerse the transducer into the appropriate disinfectant for your transducer
as shown. Follow the instructions on the disinfectant label for the duration
of transducer immersion. Do not immerse transducers longer than the
minimum time needed for your level of disinfection.
4. Using the instructions on the disinfectant label, rinse the transducer up to
the point of immersion, and then air dry or towel dry with a sterile cloth.
5. Examine the transducer for damage, such as cracks, splitting, fluid leaks, or
sharp edges or projections. If damage is evident, discontinue use of the
transducer and contact your Philips Ultrasound representative.
iE33 User Manual
310
4535 616 23901
Transducer Care
11