8Integrated Battery Handling
Battery Care
Battery care begins when you receive a new IntelliVue Cableless Measurement Device with built-in
battery and continues throughout the life of the device. The table below lists battery care activities and
when they should be performed.
Handling Precautions
Lithium ion batteries store a large amount of energy in a small package. Use caution when handling the
IntelliVue Cableless Measurement Devices with built-in battery; misuse or abuse could cause bodily
injury and/or property damage.
Handle with care.
Do not expose the device to liquids.
Do not attempt to disassemble the device.
Do not put device in autoclave.
Damaged devices should not be used anymore.
WARNING
Do not crush or puncture - mechanical abuse can lead to internal damage and internal short
circuits which may not be visible externally.
Do not incinerate the devices or expose them to temperatures above 60°C (140°F).
Activity When to perform
Perform a visual inspection Before using the IntelliVue Cableless Measurement Device
Charge the battery Upon receipt, after use, or if a low battery state is indicated.
To optimize performance, a fully (or almost fully)
discharged battery should be charged as soon as possible.
Store the device in a state of charge in
the range of 40-50% and power off the
device completely.
When not in use for an extended period of time.