Technical Alarm Messages (INOPs) 5 Patient Alarms and INOPs
75
<Pressure label>
UNPLUGGED
INOP tone.
PRESS A Pressure measurement label has been deactivated, either by
unplugging a module, or by deactivating the label in the
Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either replug the module
or reactivate the measurement label in the Measurement
Selection window.
<Pressure label>
ZERO+CHECK CAL
Numeric is replaced by -?-
PRESS Perform a zero and check the calibration of the transducer.
<pTemp label> CHECK PROBE
Numeric is replaced by -?-
INOP tone.
Predictive
Te m p
Check that the probe holder is correctly installed and that a
compatible probe is in use. If the INOP does not clear, remove
the probe from the holder then replace it. If the INOP still does
not clear, disconnect and reconnect the probe.
<pTemp label> DEACTIVATED
INOP tone.
Predictive
Te m p
The Predictive temperature has been deactivated.
<pTemp label> EQUIP MALF
Numeric is replaced by -?-
INOP tone.
Predictive
Te m p
The Predictive Temperature hardware is defective. Contact
your service personnel.
<pTemp label> INCOMPAT. Predictive
Te m p
The Predictive Temperature version is not supported by the
monitor.
<pTemp label> MEAS FAILED
Numeric is replaced by -?-
INOP tone.
Predictive
Te m p
The ambient temperature is outside the specified range (10 to
40°C /50 to 104°F). Bring the temperature into range to
continue monitoring. If the INOP does not clear, disconnect
and reconnect the probe.
<pTemp label> NO PROBE
Numeric is replaced by -?-
INOP tone.
Predictive
Te m p
Connect a probe to the Predictive temperature unit.
<pTemp label> OVERRANGE
Numeric is replaced by -?-
INOP tone.
Predictive
Te m p
The temperature at the measurement site is out of range. Check
that the probe is correctly located. If the INOP does not clear,
disconnect and reconnect the probe.
<pTemp label> PROBE MALF
Numeric is replaced by -?-
INOP tone.
Predictive
Te m p
The connected probe may be defective. Disconnect and
reconnect the probe. If the INOP does not clear, try another
probe.
pTaxil INOPS Predictive
Te m p
See <pTemp label> INOPS
pToral INOPS Predictive
Te m p
See <pTemp label> INOPS
pTrect INOPS Predictive
Te m p
See <pTemp label> INOPS
(!!)PW:Action Required Protocol
Watch
The protocol currently running requires a user response. Check
which pop-up window is displayed and provide the appropriate
response.
PW:Check Settings Protocol
Watch
Contact your service personnel. Settings could not be loaded or
interpreted correctly
PW in conflict Protocol
Watch
There is a patient information mismatch which has not yet
been resolved (>15 minutes).
INOP Message, Indication Source What to do