Technical Alarm Messages (INOPs) 5 Patient Alarms and INOPs
77
<SO
2
Label> EQUIP MALF
Numeric displays -?-.
INOP tone
SO
2
The SO
2
/SvO
2
Module or Optical Module is faulty. Unplug
and replug the Optical Module and SO
2
/SvO
2
module.
Exchange the modules. If the INOP persists, contact your
service personnel.
SO
2
INCOMPATIBLE
INOP tone
SO
2
The SO
2
Module or Optical Module is not supported. Contact
your service personnel.
<SO
2
Label> IN-VIVO CAL SO
2
The in-vivo calibration is not yet complete. Lab values must be
stored to the Optical Module to complete the calibration.
Either continue with the next steps of the current calibration or
recall the previous calibration.
<SO
2
Label>LIGHT INTENS
Numeric is replaced by -?-
INOP tone.
SO
2
The intensity changed considerably since the last light intensity
calibration. This may indicate that the catheter tip is positioned
against a blood vessel wall or that there is low blood flow.
Reposition the catheter (and perform a Light Intensity
Calibration).
<SO
2
Label> LOW LIGHT
Numeric is replaced by -?-
INOP tone.
SO
2
The optical signal levels are too low. Check that the catheter is
either in the optical reference or inserted into the patient.
Check the catheter-to-Optical Module connection. If INOP
persists, try another catheter and Optical Module.
<SO
2
Label> NO OPTMOD
Numeric is replaced by -?-
INOP tone.
SO
2
Connect the Optical Module. If the INOP persists, try another
Optical Module. Silencing this INOP switches the
measurement off.
<SO
2
Label> OPTMOD MALF SO
2
The Optical Module memory is faulty, and calibration data
cannot be stored for transport or during power failure. If this
capability is needed, use another Optical Module.
<SO
2
Label> PRE-INS CAL
Numeric displays -?-
INOP tone
SO
2
The pre-insertion calibration is running. This typically takes
one minute. During this time alarms are switched off. Wait
until the calibration is complete.
<SO
2
Label> UNPLUGGED
Numeric displays -?-.
INOP tone
SO
2
Measurement switched on and SO
2
/SvO
2
module unplugged
from the rack.
SO
2
UPGRADE
INOP tone
SO
2
The SO
2
module is currently in upgrade mode.
Monitoring is not possible in this mode.
<SO
2
Label> WARMUP
Numeric is displayed with ?
SO
2
The Optical Module has not yet reached the operating
temperature. Wait a few minutes until warm-up is finished.
SOME ECG ALRMS OFF Arrhythmia This message appears (if configured to do so) when the on/off
settings of the yellow arrhythmia alarms differ from the current
Profile.
Speaker Malfunct.
INOP tone
Monitor Contact your service personnel to check the speaker and the
connection to the speaker.
SPIRO MALFUNCTION Spirometry Module failure detected. Contact your service personnel.
SPIRO INCOMPATIBLE Spirometry Module revision not compatible with the host monitor
software revision. Contact your service personnel.
SPIRO UPGRADE Spirometry The module is running a firmware upgrade. Wait until upgrade
is completed before resuming monitoring.
SPIRO CANNOT MEAS Spirometry Measurement is at its limit, e.g. ambient pressure out of range.
INOP Message, Indication Source What to do