Software License Transfer
7-4 MX40 Repair Strategy
Required information from
defective device:
• Original Ship To Address
information (country, city)
• SW Serial Number,
OR
• HW Service Number and HW
Serial Number
(see Pg. 6-5)
Notes on the MX40 Service
Adapter and cable:
• The service adapter is for service
and demo use only.
• It is not to be used in the patient
vicinity.
• It is not to be used to power the
device when it is connected to a
patient.
• The service adapter is installed in
the battery compartment.
• The MX40 connects to the PC via
a USB port, and uses a USB cable
– but the MX40 does not operate
as a standard USB device.
• Maximum cable length = 3 ft (1m)
Note — The MX40 does not support the use of a cable longer than 3 ft.
Longer cables may result in an unacceptable drop in voltage.