Software License Transfer
7-4 MX40 Repair Strategy
Required information from
defective device:
Original Ship To Address
information (country, city)
SW Serial Number,
OR
HW Service Number and HW
Serial Number
(see Pg. 6-5)
Notes on the MX40 Service
Adapter and cable:
The service adapter is for service
and demo use only.
It is not to be used in the patient
vicinity.
It is not to be used to power the
device when it is connected to a
patient.
The service adapter is installed in
the battery compartment.
The MX40 connects to the PC via
a USB port, and uses a USB cable
but the MX40 does not operate
as a standard USB device.
Maximum cable length = 3 ft (1m)
Note The MX40 does not support the use of a cable longer than 3 ft.
Longer cables may result in an unacceptable drop in voltage.