Troubleshooting 8-13
Tele Service Batt
Note — When the above
INOP occurs, the Tele Batt
Low INOP is suppressed.
Source - MX40
Hard
The lithium-ion battery
has exceeded the
maximum charge cycle
limit and reached the end
of its useful life.
• Replace the lithium-ion
battery.
• Dispose of old battery
properly.
Tele Weak Signal
Source - MX40
Soft
• Patient is at out of
range of the radio
coverage area.
•
The
MX40 is
receiving
a weak signal for > 6
seconds
wit
h high data
loss from the AP.
• Condition may exist
for multiple devices in a
specific area
• Return patient to the
coverage area.
• If patient is in close proximity
to AP,
replace
the MX40.
Contact service.
• The AP covering the
specific area is suspect.
Contact Service.
Transmitter Off
Source - MX40
Hard RF Auto Shutoff after 10
minutes of all leads off, no
SpO
2
sensor connected,
and no short-range radio
communication.
Reattach ECG leads to patient.
Unsupported LAN
Source - MX40
Hard The MX40 (WLAN) is
connected to the Access
Point but cannot obtain
an IP address.
Correct the IP address issue.