Troubleshooting 8-13
Alarm Text
Priority
Condition
What to do
Tele Service Batt
Note — When the above
INOP occurs, the Tele Batt
Low INOP is suppressed.
Source - MX40
Hard
The lithium-ion battery
has exceeded the
maximum charge cycle
limit and reached the end
of its useful life.
Replace the lithium-ion
battery.
Dispose of old battery
properly.
Tele Weak Signal
Source - MX40
Soft
Patient is at out of
range of the radio
coverage area.
The
MX40 is
receiving
a weak signal for > 6
seconds
wit
h high data
loss from the AP.
Condition may exist
for multiple devices in a
specific area
Return patient to the
coverage area.
If patient is in close proximity
to AP,
replace
the MX40.
Contact service.
The AP covering the
specific area is suspect.
Contact Service.
Transmitter Off
Source - MX40
Hard RF Auto Shutoff after 10
minutes of all leads off, no
SpO
2
sensor connected,
and no short-range radio
communication.
Reattach ECG leads to patient.
Unsupported LAN
Source - MX40
Hard The MX40 (WLAN) is
connected to the Access
Point but cannot obtain
an IP address.
Correct the IP address issue.