Informational Messages
8-14 Troubleshooting
Informational Messages
The following table lists the Informational Text messages that may appear
in the Status Area of the MX40 display.
Informational Text Condition What to do.
Setting synchronized to
Central
The MX40 is returned to use and
settings are synchronized to reflect
any changes that may have occurred
at the Information Center.
Cleared after setting synchronization is
complete. The message displays for a
minimum of 30 seconds, depending on the
number of settings changes.
Check Revisions
The revision of the Information Center
that the MX40 is trying to connect to is
not supported.
Connect only to supported revisions of the
Information Center.
DEMO (flashing text) The MX40 is operating in Demo Mode. Cleared when the MX40 is changed to a
different operating mode. Patient monitoring is
not possible in Demo Mode. Measurements
cannot be performed, and alarms cannot be
communicated. Remove and reinsert the
battery to resume patient monitoring.
CONFIG (flashing text)
The MX40 is operating in
Configuration Mode.
Cleared when the MX40 is changed to a
different operating mode.
Incompatible Option Masimo SET or FAST SpO2 device
configured with wrong option.
Cleared automatically after temporary display.
SERVICE (flashing text)
The MX40 is operating in Service
Mode.
Cleared when the MX40 is changed to a
different operating mode. Patient monitoring is
not possible in Service Mode. Measurements
cannot be performed, and alarms cannot be
communicated. Remove and reinsert the
battery to resume patient monitoring.
SRR Started
Short-range radio is powered on (used
with Cableless Measurement Devices
and when paired to a patient monitor).
Cleared automatically.
No OR-ECG leadset
connected
IntelliVue Patient Monitor leadset
adapter is in use. This message is a
reminder not to use OR-ECG leadsets
with the adapter.
Continues to flash as a reminder while the
leadset adapter is in use. Cleared when
adapter is removed from MX40.
Resp Avail User confirms that an OR-ECG
leadset is not attached with the
adapter.
Cleared automatically after one minute.
Resp not available The Resp feature is not enabled. Software option must be installed to enable
Resp.
SpO2 not available SpO2 sensor is connected to MX40
but the device does not have the
SpO2 option enabled/available.
Enable MX40 software option/upgrade
hardware.
Switch to Masimo SET
Leadset
Incorrect leadset is connected to
Masimo device.
Cleared automatically after connecting
Masimo leadset.