Informational Messages
Troubleshooting 8-15
Informational Text Condition What to do.
Switch to Philips FAST
Leadset
Incorrect leadset is connected to
Philips FAST device.
Cleared automatically after connecting Philips
FAST leadset.
SRR Network Scan Short-range radio channel scan in
progress.
Cleared automatically when complete.
SRR Channel <chan
num>
Short-range radio channel scan is
complete and best channel is
selected.
Cleared automatically after selection.
SRR Searching Sensor
Short-range radio is attempting to
associate with Cableless
Measurement Devices.
Cleared automatically.
SRR Stopped Short-range radio is in power saving
mode and no longer searching for
Cableless Measurement Device until
the Add/Remove SmartKey is touched
again
Cleared automatically or when the
Add/Remove SmartKey is touched.
cL NBP Assigned Short-range radio is assigned to an
NBP Cableless Measurement Device.
Cleared automatically after one minute.
cL SpO
2
Assigned
Short-range radio is assigned to an
SpO
2
Cableless Measurement Device.
Cleared automatically after one minute.
SRR Searching Monitor Short-range radio is attempting to
associate with a patient monitor.
Cleared automatically.
SRR Assoc Monitor Short-range radio is connected to a
patient monitor.
Cleared automatically.
SRR Link Test
Short-range radio link testing is in
progress before final association.
Cleared automatically. Icon replaces
message.
SRR Unavailable w/ Resp Respiration monitoring is enabled,
therefore the Short-range radio is
disabled.
Turn the Resp parameter off at the
Information Center.