Additional Hardware Troubleshooting Information
8-16 Troubleshooting
Additional Hardware Troubleshooting Information
Problem Condition What to do.
The following icon is displayed
on the MX40:
865352 802.11 MX40 does not operate, just
displays this icon.
• The 865352 MX40 cannot use AA
batteries. It will only operate with
the Philips MX40 Li-ion battery.
Remove the AA batteries and battery
adapter and replace with a Philips
MX40 Li-ion battery.
• There is damage to a battery contact
in the MX40 battery compartment,
and the MX40 cannot recognize that
a Philips MX40 Li-ion battery is
installed. The MX40 will require
replacement via the Exchange
Process, or for USA and Canada
only, repair (at the Philips Lifecycle
Support Center).
• Failure of the Philips Li-ion battery.
Replace the battery.