Possible User Interface Issues
8-22 Troubleshooting
Possible User Interface Issues
The MX40 display does not turn on.
The AA Battery Tray may be inserted backwards.
The user may not understand that they need to touch the blue
Main Screen button for two seconds.
The MX40 display is on but does not react to touch.
The screen is locked and needs to be unlocked using the Unlock
SmartKey.
a. Select the Smartkey button.
b. Scroll to the second page.
c. Select the Lock/Unlock Smartkey.