Electromagnetic Compatibility
Safety Standards & Specifications 10-11
Electromagnetic Compatibility (EMC) Specifications
Take special precautions regarding electromagnetic compatibility (EMC)
when using medical electrical equipment. You must operate your monitoring
equipment according to the EMC information provided in this book. Portable
and mobile radiofrequency (RF) communications equipment can affect
medical electrical equipment.
Accessories Compliant with EMC Standards
All accessories listed in the accessories section comply, in combination with
the MX40, with the requirements of IEC 60601-1-2.
Warning
Using accessories other than those specified may result in increased
electromagnetic emission or decreased electromagnetic immunity of the
monitoring equipment.
Electromagnetic Emissions
Emissions Test Compliance
Avoiding Electromagnetic
Interference
Radio Frequency (RF)
emissions
Group 1 TheMX40 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions CISPR
11
Class B
The MX40 is suitable for use in all
establishments.
Harmonized emissions Not Applicable Device is battery powered only
Voltage
fluctuations/Flicker
emissions
IEC 61000-3-3
Not Applicable