Battery Specifications
Safety Standards & Specifications 10-15
Battery Specifications
Battery Life
Note —Battery life specifications are stated with a 95% confidence level that
the average run time will be at least this long assuming a random sample of
the population with a student-t distribution.
The battery life specifications listed below are based on the use of three
Duracell MN 1500 batteries. Battery life for other brands may differ.
Telemetry Mode
Networked
(Display Off)
Battery Life
(1.4GHz
p/n 865350)
Battery Life
(2.4GHz
p/n 865351)
ECG Only 37.0 hours 33.0 hours
ECG/SpO
2
FAST
Continuous (ECG operation
available.)
22.0 hours
(time to No Signal
INOP)
20.0 hours
(time to No Signal
INOP)
ECG/SpO
2
FAST
Continuous (Continuous
SpO2 operation available.)
10.0 hours
(time to No SpO2T,
Batt Low INOP)
10.0 hours
(time to No SpO2T,
Batt Low INOP)
ECG/SpO
2
Masimo
Continuous (ECG operation
available.)
20.0 hours
(time to No Signal
INOP)
18.0 hours
(time to No Signal
INOP)
ECG/SpO
2
Masimo
Continuous (Continuous
SpO2 operation available.)
9.0 hours
(time to No SpO2T,
Batt Low INOP)
8.0 hours
(time to No SpO2T,
Batt Low INOP)
ECG/SpO2 FAST Auto (5
min.) (ECG operation
available.)
35.0 hours
(time to No Signal
INOP)
30.0 hours
(time to No Signal
INOP)
ECG/SpO2 FAST Auto (5
min.) (SpO2 operation
available)
18.0 hours
(time to No SpO2T,
Batt Low INOP)
16.0 hours
(time to No SpO2T,
Batt Low INOP)
ECG/SpO2 Masimo Auto (5
min.) (ECG operation
available.)
30.0 hours
(time to No Signal
INOP)
28.0 hours
(time to No Signal
INOP)
ECG/SpO2 Masimo Auto (5
min.) (SpO2 operation
available)
15.0 hours
(time to No SpO2T,
Batt Low INOP)
14.0 hours
(time to No SpO2T,
Batt Low INOP)
ECG/SpO
2
Manual (FAST
and Masimo)
In this mode battery life is dependent on the
usage rate and will range between the ECG
Only battery life and the ECG/SpO
2
Continuous battery life.