Measurement Specifications
10-30 Safety Standards & Specifications
Distributed Alarm System Delay Specifications
For complete Distributed Alarm System Delay information and specifications,
see the Patient Information Center iX Instructions for Use.
Respiration
Parameter Specification
Leads Used for Measurement RA, LL (standard) or I, A (EASI)
Range Adult/Pedi: 0 to 120 rpm
Bandwidth 0.3Hz to 2.5Hz (-6dB)
Noise Less than 25 mOhm (rms) referred to the
input
Calibration Signal
Signal: 1 Ohm p-p; Accuracy:
±20%
Respiration Rate Resolution 1 rpm
Respiration Accuracy
±1 rpm for 0-120 rmp
Auxiliary Current, Respiration Excitation
Signal
< 470 uA rms @48KHz, sinusoidal waveform
Respiration Alarm
Alarm Range Delay
High Adult/Pediatric: 10 to 100 rpm ≤ 15 seconds
Low Adult/Pediatric: 0 to 95 rpm for limits from 0 to 20 rpm: max. 4 seconds
for limits above 20 rpm: max. 15 seconds
Apnea Alarm 10 to 40 seconds Incremental delay 5 seconds max.
FAST SpO
2
Parameter Specification
SpO
2
Measurement
Range (Calibration and
Display)
0 to 100%
SpO
2
Accuracy
See “Performance Specification Details for Philips FAST Sensors” MX40
C.01 Instructions for Use, Chapters 9 and 15..