Measurement Specifications
10-32 Safety Standards & Specifications
Masimo SET SpO
2
Parameter Specification
SpO
2
Measurement
Range (Calibration and
Display)
0 to 100%
SpO
2
Accuracy See “Performance Specification Details for Masimo SET Sensors” MX40 C.01
Instructions for Use, Chapters 9 and 15..
SpO
2
Resolution 1%
SpO
2
Numerics -
Averaging
5 - 20 seconds (default = 10 seconds)
SpO
2
& Pulse Numerics -
Update Rate
Typical numeric update rate is once per second..
Pleth Wave- Sampling
Rate
125 sps
Technical Alarms
(INOPs)
Triggered if the sensor is disconnected, if a pulse is not detected, if the signal
is noisy, if light interference is detected, if the sensor is defective, if the
measurement is erratic, or if the module is malfunctioning
Wavelength Range 500 to 1000 nm
NoteInformation about wavelength range can be especially useful to
clinicians (e.g., clinicians performing photodynamic therapy).
Pulse Rate Measurement Range: 25 to 240 bpm
Accuracy with no motion: 3 bpm
Accuracy with motion: 5 bpm
Resolution: 1 bpm
Display of SpO
2
Numerics
SpO
2
values are displayed as xxx % SpO
2
to meet ISO 80601-2-61
Emitted Light Energy < 15 mW
SpO
2
Accuracy Specifications
For complete SpO2 Accuracy Specification information, see the IntelliVue
MX40 C.01 Instructions for Use, Chapters 9 and 15.