MX40 Test & Inspection Matrix
3-2 Test and Inspection
MX40 Test & Inspection Matrix
Test Block
Name
Test or "Inspection" to Perform
What to Record
on Service
Record
Visual Test: Inspect the MX40 (and packing material if applicable) for
obvious signs of damage. Also check external leads and
accessories.
Expected Test Results: The device does not have any obvious
signs of damage = Pass
V:P or
V:F
where P=Pass
F=Fail
Power On:
Remove patient cable. Insert battery into the MX40. The MX40
will go through its self-test and pass. Make sure that an ECG
wave appears on the screen and the battery gauge displays
battery status. Check the INOP Area for any equipment
malfunctions.
The expected test result is pass: the MX40 boots up and
displays an ECG wave and the battery gauge displays battery
status. The wave will be a flat line if no simulator is attached.
Expected Test Results: Expected answer is "yes". If so, Power
On test is passed.
PO:P or
PO:F
where P=Pass
F=Fail
Performance:
1. Insert battery into the MX40 for the channel being tested.
2. Attach a patient cable to the MX40 and an ECG simulator.
3. At the Information Center assign the MX40 being tested to a
Sector. Ensure that the Multi-Function Button is turned "on",
and turn on the SpO
2
parameter if the MX40 being tested has
the SpO
2
option. Set the mode to Continuous.
4. An ECG waveform should be visible at the Information
Center.
5. If the MX40 has the SpO
2
option, connect an SpO
2
sensor
and apply the SpO
2
sensor to yourself. Confirm that the
MX40 completes a successful measurement.
P:P or
P:F
where P=Pass
F=Fail
6 .Set the SpO
2
mode to the customer's desired setting,
Continuous or Spot Check/Manual.
7. Place the device in Standby. At the Information Center,
resume monitoring.
8. Press the Multi-Function Button on the MX40. The button
press should generate one of the following, depending on
the configured setting:
• Nurse Call & Record - Nurse Call alarm and a recording
generated at the Information Center.
• Nurse Call Only - Nurse Call alarm at the Information Center.
• Record Only - A recording generated at the Information
Center.