Monitoring Mode
4-2 Operating Modes
Monitoring Mode
Monitoring Mode is the normal operating mode of the MX40 and a
password is not required.
Warning
If the MX40 displays a SW License Required message, remove the device
from service. The device must be repaired/replaced.
Caution
Do not use pneumatic tube systems to transport this device or the
rechargeable lithium-ion battery. Damage may result.
Telemetry Mode Use
To minimize patient disruption, the MX40 operates in Telemetry Mode
when connected to the Information Center. In Telemetry Mode, the local
volume is set to zero and the display is off. You can activate the display at
any time by touching the Main Screen button for two seconds. All active
alarms can be viewed when the display is on, however audible alarm
indicators are not annunciated. Regardless of the display status, all
measurement data is being sent to the Information Center. Telemetry Mode
is only available when connected to the Information Center.
Monitor Mode Use
In Monitor Mode, the display is always on for easy viewing, and should an
alarm condition occur, it will be announced locally at the MX40 and at the
Information Center if networked connected. Be aware that when the
display is on and alarms are audible, battery consumption will increase.
To use Monitor Mode:
1. Press the SmartKeys Button.
2. Press the Mode: Telemtry / Mode: Monitor SmartKey and choose
Mode: Monitor.