Monitoring Mode
Operating Modes 4-5
Multi-Function Button
Button Function
Depending on configuration at the Information Center:
• generates a Nurse Call;
• Initiates a Delayed Recording;
• Both, or;
• None
Note — If appropriate, patients should be informed that
the Multi-function button on the MX40 is not the primary
method for generating a Nurse Call.
Silence/Acknowledge Alarm Button
Button Function
• Initiates a local silence/acknowledgment of all
active alarms when enabled (IIC).
• Initiates a global silence/acknowledgment of all
active alarms (IIC iX).
• Silences the "Find Device" sound.
Note — Alarms at the MX40 can be silenced
from the Information Center. When silenced from
the Information Center, the alarm sound is not
silenced at the Information Center until it receives
feedback from the MX40. This may take several
seconds.
SmartKeys Button
Button Function
Displays the SmartKey Menu on the touch screen.
Main Screen Button
Button Function
• Activates the display screen if touched for two
seconds.
• Cycles through the display screens if touched
repeatedly.
• Resumes from Standby.
• When pressed from a sub-menu, returns display to the
Main Screen.