Monitoring Mode
4-10 Operating Modes
Check that the above actions occur and that you are connected to the
Information Center when in the wireless coverage area. If you are not in
range of the Information Center, the device continues to sound and the
display remains active.
A successful power-on self test will then transition the MX40 to the start-up
screen. Selectable background colors can be configured and display on the
screen for assistance with device identification. This can be helpful when
devices are in a pooled use setting.
You must visually check that a waveform is present on the display. You can
access further status information is by touching the status area on the
display.
On a daily basis, the clinician should inspect the MX40 and accessories.
Replace any damaged equipment or accessories.
Touch Sensitive Display
The display area of the MX40 is touch sensitive and can be touched with
your fingertip or thumb through the Philips Waterproof Carrying Pouch or
medical gloves. Do not use a pen or stylus. When touched, the element is
highlighted. After removing your fingertip or thumb, a white dot is
displayed to guide you by showing the location of your touch. When the
display is locked, a red dot is displayed.
Navigating
Touching the Navigation Bar on the right of the display will scroll through
additional display items. Solid downward arrows indicate there are
additional elements that are not currently displayed. The arrows briefly
illuminate when touched. Your selection from the menu also illuminates
when touched.
Selecting Display Elements
Touch a display element to get to the actions linked to that element. For
example, touch the Patient Information element to call up the Patient Info
window, or touch the HR numeric to call up the Setup ECG menu. Touch
the ECG waveform to call up the wave selection menu.