Monitoring Mode
4-16 Operating Modes
Note — Primary or secondary waveform configuration changes made at the
Information Center change what is displayed on the MX40. But changes
made locally at the MX40 do not affect settings at the Information Center.
Selecting Pleth or RESP to display at the Information Center affects ECG
wave storage for future review at the Information Center.
Battery Information
Battery Safety Information
Warnings
• The battery compartment door must be closed during defibrillation.
• Use the Philips Rechargeable Lithium-ion Battery or 3 Duracell Alkaline
batteries, size AA,
MN 1500, 1.5V, to ensure specified performance and
correct battery gauge reporting.
• Outdated, mismatched, or poor-quality batteries can give unacceptable
performance (e.g., insufficient Battery-Low warning time). If you are
using disposable batteries, the use of fresh high-quality alkaline
batteries is strongly recommended.
• Certain failure conditions, such as short circuits, can cause a battery to
overheat during use. High temperatures can cause burns to the patient
and/or user. If the MX40 becomes hot to the touch, remove it from the
patient and place it aside until it cools. Then remove the batteries and
discard them.
• If you receive a TELE BATTERY LOW, TELE BATT EMPTY, REPLACE
TELE BATT, or TELE BATTERY TEMP alarm, the batteries must be
promptly replaced. If these conditions are not corrected, they will result
in a device shutdown and cessation of monitoring. These INOPS also
apply when monitoring with Cableless Measurements.
• Disposable batteries should be removed from the MX40 at the end of
the battery’s useful life to prevent leakage.
• Use only the Philips manufactured Rechargeable Lithium-ion Battery.
• Do not use other types of rechargeable batteries. Other types of
rechargeable batteries will adversely affect:
- Battery gauge performance
- Battery low warnings
- Battery life performance